Recalibration

HTL pipettes are calibrated by gravimetric method, using distilled water, at the temperature 20±1°C, according to EN ISO 8655 standard. If during pipette operation you find that the accuracy error (the difference between the real aspirated volume and the preset volume) exceeds the permissible value given in the Instruction Manual, the pipette recalibration procedure should be carried out. Before starting the recalibration it is necessary to check whether the following requirements have been fulfilled during error determination:

  • the ambient temperature, and the temperature of the pipette, tips and water was identical
  • the density of the liquid used is close to that of distilled water
  • a balance with appropriate sensitivity has been used
Volume tested
(µl)
Measurement Accuracy
(mg)
Standard Deviation
(mg)
≥ 0,1-10 ≤ 0,001 ≤ 0,002
≥ 10-100 ≤ 0,01 ≤ 0,02
≥ 100-1000 ≤ 0,1 ≤ 0,2
≥ 1000 ≤ 0,1 ≤ 0,2
  • mg/µl conversion factor has been taken into account
  • the requirements given in sections Aspirating and Recommendations of Instruction Manual have been fulfilled.

If the above conditions are satisfied and the accuracy error for selected volume exceeds the permissible value - see in Instruction Manual, the pipette recalibration procedure should be carried out.

The recalibration can be performed within one full turn of the key to the right or to the left only.

Recalibration conditions

  • Ambient temperature and the temperature of the pipette, tips and liquid should be within the range 20-25°C and stabilized during weighing within ±0.5°C
  • Measurements should be conducted using distilled water
  • Balance sensitivity should be suitable for the volume to be controlled.


Step by step recalibration procedure