Maintenance
HTL Pipettes do not require any special maintenance
procedures. The clear technical design in combination with
the proper choice of durable materials such as PVDF make HTL
pipettes and pipetors long lasting instruments. HTL aims to
provide the users with easy to handle pipettes of the
highest safety standards. If the user follows the
Instruction
Manual and takes
care of cleaning the pipette after working with aggressive
materials, seeds and reagents it will function perfectly for
years.
Here are some basic instructions to follow, which will make
your pipette a safe and precise tool for your day by day
activity:
- Do not use pipette without a tip.
- Do not invert the pipette while the tip is filled
with liquid or wet.
- Always keep the pipettes in the stand after use.
Following the above recommendations will make cleaning
and sterilization a regular, easy to do action rather than
extraordinary, tiring and time-consuming effort.
Cleaning
External surfaces of the handgrip, the pipetting pushbutton, the
ejector pushbutton, the shaft nut and the
adjustment knob may be cleaned using a cloth dampened in
isopropyl alcohol. The remaining parts removed from the
pipette during pipette disassembly may be washed with
distilled water or isopropyl alcohol
Sterilization
Applies to DISCOVERY+ Series Pipettes and CLINIPET+:
The pipette can be sterilized in the autoclave at 121°C for
20 minutes. After sterilization, the pipette should be dried
and cooled to room temperature
It is recommended:
- to sterilize the pipettes in autoclave with an initial
vacuum and drying cycle,
- prior to sterilization unscrew the shaft nut slightly
in the DV2-DV1000 pipettes, and unscrew the shaft
slightly in the DV5000 and DV10000. After autoclaving
these parts should be screwed tight again. The precision
of the results should not alter if the pipetting process
and autoclaving are carried out as described in this
manual. Because a slight change in the accuracy of the
dosage may occur, it is recommended to:
- check the calibration of the pipette after
every 10 autoclaving cycles.
Applies to LAB MATE+ pipettes
and V3 SERIES pipettes:
The shaft, shaft nut, ejector, piston assembly, seal and
O-ring may be autoclaved at a temperature of 121°C for 20
minutes at pressure 0.1 MPa. The shaft of the 5000 and 10000
models should be autoclaved without the filter. The handle
and the outside of the pipette may be cleaned using a cloth
damped in isopropanol.
Warning
Replacing the piston requires recalibration of the
pipette. We recommend that this work be conducted by HTL
Authorized Service.
Before returning the pipette for service please make sure it
is free from any chemicals, radioactive and microbiological
contamination otherwise this might affect the security of
people involved in transportation or service.
|